l
PCT performs a percent change operation. The result is the percent change of the measurement from
the reference value:
Result = ((Measurement – Reference) / Reference) * 100
For the PCT function, the reference value can be automatically selected using the first measurement
as the reference value (see
CALCulate:SCALe:REFerence:AUTO
), or it can be specified by
. % scaling applies to all measurement functions except Ratio, Continuity, and
Diode.
l
SCALe performs an Mx-B operation. The result is the measurement multiplied by the gain value M
(
). Mx-B scaling applies to
all measurement functions except Ratio, Continuity, and Diode.
l
Scaling function results must be in the range of -1.0E+24 to -1.0E-24, or +1.0E-24 to 1.0E+24. Res-
ults outside these limits will be replaced with -9.9E37 (negative infinity), 0, or 9.9E37 (positive infinity).
l
You must execute CALCulate:SCALe:FUNCtion
before
executing CALCulate:SCALe[:STATe], otherwise
an incompatibility error will be generated.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
l
Scaling is set to OFF when you change measurement functions (for example, changing from DCV to
ACV). You must re-enable scaling after changing measurement functions.
CALCulate:SCALe:GAIN {<
gain
>|MIN|MAX|DEF}
CALCulate:SCALe:GAIN? [{MIN|MAX|DEF}]
Stores a gain value
M
for the scaling function in
Parameter
Typical Return
-1.0E+15 to -1.0E-15, 0.0, +1.0E-15 to 1.0E+15. Default: 1.0. +1.0002
Enable
Mx-B
scaling with a gain of 100 and an offset of 5.
CALC:SCAL:FUNC SCAL
CALC:SCAL:GAIN 100
CALC:SCAL:OFFS 5
CALC:SCAL:STAT ON
l
Applies only to the 34465A and 34470A.
l
The instrument sets the gain value to 0.0 after a Factory Reset (
) or Instrument Preset
(
).
SCPI Programming Reference
224
Keysight True
volt
Series Operating and Service Guide
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...