UNIT:TEMPerature {C|F|K}
UNIT:TEMPerature?
Selects the units (°C, °F or Kelvin) to be used for all temperature measurements except the
query.
Parameter
Typical Return
{C|F|K}. Default: C.
C, F, or K
Make a four-wire RTD measurement and return the result in °F:
UNIT:TEMP F
MEAS:TEMP? FRTD
Typical Response: +6.8201
l
The command also accepts CEL or FAR, but the query returns C or F.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
See Also
Keysight True
volt
Series Operating and Service Guide
209
SCPI Programming Reference