62
Chapter 4 Calibration Procedures
Internal Timebase Verification
4
Internal Timebase Verification
This test verifies the output frequency accuracy of the instrument. All
output frequencies are derived from a single generated frequency.
1
Connect a frequency counter as shown below (the frequency counter
input should be terminated at 50
Ω
).
2
Set the instrument to the output described in the table below and
measure the output frequency. Be sure the instrument output is enabled.
*
The error is ± 100 Hz within 90 days of calibration, or ± 200 Hz within one year.
3
Compare the measured frequency to the test limits shown in the table.
Keysight
33210A
Measurement
Function
Amplitude
Frequency
Nominal
Error
Q
Sine Wave
1.00 Vpp
10.000,000,0 MHz
10.000 MHz
± 200 Hz
*