39
Chapter 3 Front-Panel Menu Operation
To Select the Output Termination
4
3
To Select the Output Termination
The
Keysight
33210A has a fixed series output impedance of 50 ohms
to the front-panel Output connector. If the actual load impedance is
different than the value specified, the displayed amplitude and offset
levels will be incorrect. The load impedance setting is simply provided
as
a convenience to ensure that the displayed voltage matches the
expected
load.
1 Press
.
2 Navigate the menu to set the output termination.
Press the
Output Setup
softkey and then select the
Load
softkey.
3 Select the desired output termination.
Use the knob or numeric keypad to select the desired load impedance or
press the
Load
softkey again to choose “High Z”.
To Reset the Function Generator
To reset the instrument to its factory default state, press
and then
select the
Set to Defaults
softkey. Press
YES
to confirm the operation.
For a complete listing of the instrument’s power-on and reset conditions,
see “
Keysight
33210A Factory Default Settings” in the User’s Guide.
Содержание 33210A
Страница 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator Service Guide ...
Страница 2: ......
Страница 3: ......
Страница 10: ...8 ...
Страница 13: ...1 1 Specifications ...
Страница 19: ...2 2 Quick Start ...
Страница 37: ...3 3 Front Panel Menu Operation ...
Страница 52: ...50 3 ...
Страница 53: ...4 4 Calibration Procedures ...
Страница 96: ...94 Chapter 4 Calibration Procedures Calibration Errors 4 ...
Страница 97: ...5 5 Block Diagram ...
Страница 101: ...99 Chapter 5 Block Diagram Block Diagram 4 5 ...
Страница 103: ...6 6 Disassembly and Repair ...
Страница 128: ...126 Chapter 6 Disassembly and Repair Replaceable Parts 6 ...
Страница 129: ...7 7 Backdating ...