21
Chapter 2 Quick Start
To Set the Output Frequency
4
2
To Set the Output Frequency
At power-on, the function generator outputs a sine wave at 1 kHz with
an amplitude of 100 mV peak-to-peak (into a 50
Ω
termination).
The following steps show you how to change the frequency to 1.2 MHz.
1 Press the
“
Freq
”
softkey.
The displayed frequency is either the power-on value or the frequency
previously selected. When you change functions, the same frequency is
used if the present value is valid for the new function. To set the
waveform period instead, press the
Freq
softkey again to toggle to the
Period
softkey (the current selection is highlighted).
2 Enter the magnitude of the desired frequency.
Using the numeric keypad, enter the value “1.2”.
3 Select the desired units.
Press the softkey that corresponds to the desired units. When you select
the units, the function generator outputs a waveform with the displayed
frequency (if the output is enabled). For this example, press
MHz
.
Note:
You can also enter the desired value using the knob and cursor
keys.
Содержание 33210A
Страница 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator Service Guide ...
Страница 2: ......
Страница 3: ......
Страница 10: ...8 ...
Страница 13: ...1 1 Specifications ...
Страница 19: ...2 2 Quick Start ...
Страница 37: ...3 3 Front Panel Menu Operation ...
Страница 52: ...50 3 ...
Страница 53: ...4 4 Calibration Procedures ...
Страница 96: ...94 Chapter 4 Calibration Procedures Calibration Errors 4 ...
Страница 97: ...5 5 Block Diagram ...
Страница 101: ...99 Chapter 5 Block Diagram Block Diagram 4 5 ...
Страница 103: ...6 6 Disassembly and Repair ...
Страница 128: ...126 Chapter 6 Disassembly and Repair Replaceable Parts 6 ...
Страница 129: ...7 7 Backdating ...