General Power Meter Functions
94
Keysight N1913/1914A User’s Guide
Max Hold/Min Hold
The max hold/min hold setting on the measurement setup window can be set on
the front panel or via SCPI.
Min hold is the minimum of all measurements since the start time. The min hold
reading will be updated as new minimum reading appears.
Max hold is the maximum of all measurements since the start time. The max
reading will be updated as new maximum reading appears.
The max hold/min hold reading can be resettable by pressing the
key.
Figure 2-31
Max hold/min hold measurement is performed on the ‘HOLD’
block
SENSe1:
Input from
SENSe1
block
SENSe2:
Input from
SENSe2
block
(N1914A only)
:FEED
FEED1
FEED2
:MATH
“A”|”B”
“A-A”|“A/A“
“B-A”|“B/A“
“A-B”|“A/B“
“B-B”|“B/B“
A
B
:GAIN
:HOLD
:REL
CALCulate Block
AVG
AVG
Max Hold/
Min Hold