General Power Meter Functions
102
Keysight N1913/1914A User’s Guide
Zeroing and Calibrating the Power Meter
This section describes how to zero and calibrate the power meter. You should
always zero the power meter prior to calibrating it.
Zeroing the Power Meter
Zeroing adjusts the power meter for a zero power reading with no power applied
to the power sensor. During zeroing, which takes approximately 10 seconds, the
wait symbol is displayed.
Figure 2-41
Zeroing pop-up
When to Zero?
Zeroing of the power meter is recommended:
– When a 5 oC change in temperature occurs.
– Every 24 hours.
– Prior to measuring low level signals. For example, 10 dB above the lowest
specified power for your power sensor.
To zero the power meter:
– Press
and the channel
softkey.
– The
Zeroing
pop-up is displayed.
Zero
NOTE
On dual channel meters you can zero each channel independently or both
channels sequentially by pressing the
softkey - choosing
,
or
Zero
Zero A Zero B
Zero Both
Содержание 1914A
Страница 1: ...Keysight N1913 1914A EPM Series Power Meters User s Guide...
Страница 24: ...24 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 54: ...1 Introduction 54 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 206: ...7 Using U2000 Series USB Power Sensors 206 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 240: ...9 Maintenance 240 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 242: ...10 Characteristics and Specifications 242 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...