U3047AM08
Hardware Lock-link Installation (U3021-60002)
Hardware Lock-link Installation (U3021-60002)
If your system is to be rack mounted, this installation procedure does not have
to be performed.
Locking the Test Set to the Network Analyzer
1. The lock-link kit (U3021-60002) includes:
— 5023-0132 − Lock-links kit (left, right pair and screws), analyzer
— 0515-2317 − Screw T15 M3.5 x 0.6 12 mm
— N5242-20138 − Right Lock-link, test set
— N5242-20139 − Left Lock-link, test set
2. Remove the feet from the bottom of the analyzer.
3. Remove the 2 lower standoffs from the rear panel on the analyzer using a
T20 Torx driver.
Figure 5
Rear and Bottom Feet
The opening of covers or removal of parts is likely to expose dangerous
voltages. Disconnect the instrument from all voltage sources while it is
being opened.
Standoffs (x2)
Feet (x4)
18
Keysight U3047AM08 User's and Service Guide