U3042AM08
Cal Kit Operational Check
1-Port Calibration and Verification Procedure
1.
Connect the ECal or the mechanical cal kit to Port 1 or the port you are testing.
Torque to 8 in-lb. For further information refer to the Help menu.
2.
Perform a 1-Port Calibration on Port 1. On the analyzer, select
Response
>
Cal
>
Start Cal
>
Calibration Wizard
-OR-
(Response > Cal > Other Cals)
.
a.
If using a mechanical cal kit, select
SmartCal (Guided Calibration)
>
Next
.
b.
If using an ECal module, connect the ECal to a PNA USB port. Select
Use
Electronic Calibration (ECal)
>
Next
.
3.
Continue following the Cal Wizard prompts. In the Select Calibration Ports and
ECal Module dialog box click
Clear All
, then select
Port 1
>
Next
>
Measure
.
4.
Ensure the Cal Kit you are using is indicated on the right side of the window.
Figure
50
1-Port Calibration
5. Continue to follow the prompts until the "Calibration Completed" dialog box
appears.
6. At the Calibration Completed prompt, select
Save As User Calset
, type the name
999.1
. Overwrite the Calset if it already exists and
Save
.
NOTE
If you do not have a key board, select
Save As User Calset
>
Edit Name
and save as
999.x
.
X is the port number you are calibrating. Use the numeric keypad on the
analyzer front panel to enter “999.1.”
Figure
51
Calibration Complete
68
Keysight U3042AM08 User's and Service Guide