Example 1
Test Set Port 7 (Tx & Rx) and Test Set Port 16 (Tx & RX)
In this example we’ll combine test set ports 7 and 16 data sets to be able to transmit (Tx) and
receive (Rx) without Ext Noise Source:
Looking at
"Combined Tx and Rx Port Selections"
entries:
•
Port 7:
0.85 16.0 32.17
•
Port 16: 64.32
Port Address
Data
7
0,16,32
85,0,17
Port Address
Data
16
64
34
SCPI Command Entry Method
CONT:EXT:TEST:DATA
0,85; delay 20ms
CONT:EXT:TEST:DATA
16,0; delay 20ms
CONT:EXT:TEST:DATA
32,17; delay 20ms
CONT:EXT:TEST:DATA
64,34; delay 20ms
To add the Ext Noise Source path to port 7, change "32,17" to
"32,145" as shown in
U3022AM42
Address and Data Values
Keysight U3022AM42 User's and Service Guide
47