U3022AM42
Controlling the Test Set with N5242A/B
RF Path Configuration
for Analyzer with Noise Figure
Capability
If the analyzer has noise figure capability
,
verify that the path configuration is set to
Default. If it is not in Default configuration, Port 5 will not operate correctly.
1. Select
Trace/Chan > Channel > Hardware Setup > Path Cong...
-OR-
(Instrument >
Setup >
Internal Hardware > RF Path Config...)
and in the drop-down
menu select
Default
>
OK
. Save this configuration as a User Preset by selecting
Save
>
User Preset > Save current state as User Preset
.
Do not use the factory Preset (User Preset Off), the analyzer will return to Option 029
path configuration.
2. Verify that the Port 1 Noise Tuner Switch is set to external. Select
Utilities System
>
Configure
>
Preferences
-OR-
(Utility > System > System Setup > Preferences > User
Preset)
.
If not, select
Meas: Port 1 Noise Tuner Switch is set to external
.
Figure
3
0
RF Path Configuration
Keysight U3022AM42 User's and Service Guide
35