
U3022AM06
Controlling the Test Set
Figure 50
Command Console for 'B' Model Analyzers
SCPI Command Processor Console
There are two methods in which the PNA-X Command Processor Console can be used to
control the Test Set internal switches to configure the Multiport system for S-Parameter
measurements. When the Command Processor Console is opened as instructed
previously, examples of the two control methods will be provided.
Method 1 - Using GPIB/SCPI Command Values
This method is available while the system is in Multiport mode only. The Test Set internal
switch settings are programmed into the Test Set firmware; they are configured by the PNA-X
S-parameter values.
Method 1 provides more PNA-X control capability than Method 2.
The following example shows two executable commands needed to create an S-parameter
measurement on your Multiport system.
Figure 51
Method 1 - Using GPIB/SCPI Command Values
58
Keysight U3022AM06 User's and Service Guide
Содержание U3022AM06
Страница 6: ...Table of Contents...
Страница 7: ...U3022AM06 Keysight U3022AM06 User s and Service Guide 7...
Страница 80: ...Figure 66 U3022AM06 System Block Diagram U3022AM06 Option 129 80 Keysight U3022AM06 User s and Service Guide...
Страница 101: ...Keysight U3022AM06 User s and Service Guide 101...