![Keysight Technologies U3022AM06 Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/keysight-technologies/u3022am06/u3022am06_users-and-service-manual_1962380040.webp)
U3022AM06
System Operational Checks
System Operational Checks
The following procedure will confirm that the RF interface cables between the Test
Set and Analyzer are installed and the system is working correctly.
Before beginning this procedure you should complete the following steps:
•
Stack the Analyzer on top of the Test Set.
•
Install the Test Set I/O cable on the rear panel.
•
Install all of the RF interface cables on the front panel.
•
Install the female 3.5 mm short (85052-60007) to Port 1 on the Analyzer.
Equipment Setup
•
Turn
on
the Test Set and Network Analyzer.
•
Preset the Analyzer and complete the following settings:
— Set to Multi-Port Mode (refer to
“Multiport Mode (Option 551)” on page
— Frequency range: 10 MHz to 26.5 GHz or highest frequency
— Set the Analyzer to measure S11
Verify Results
The trace ripple (peak-peak variation) will be higher than when using an ECal Module
due to variation in your Short’s performance. If response is in question perform
Kit Operational Check” on page
Table 13
The 3.5 mm short will be relocated to each Test Port and an uncorrected Sxx
reflection response trace should be displayed that resembles the following plots
indicated in the table below. If you suspect an RF signal path problem, refer to the RF
cable path diagram and verify the I/O cables associated with this signal path for
proper installation.
Reflection Response Results
Reflection Port
Response Trace
Cable Path Diagram
Ports 1-2
Ports 3-8
40
Keysight U3022AM06 User's and Service Guide
Содержание U3022AM06
Страница 6: ...Table of Contents...
Страница 7: ...U3022AM06 Keysight U3022AM06 User s and Service Guide 7...
Страница 80: ...Figure 66 U3022AM06 System Block Diagram U3022AM06 Option 129 80 Keysight U3022AM06 User s and Service Guide...
Страница 101: ...Keysight U3022AM06 User s and Service Guide 101...