Measurement Setup
Figure
5
Windows Start Menu
1.
connect the two RF cable ends together using a 3.5 mm
female-to-female adapter.
2.
Normalize the displayed response trace on the network analyzer.
3.
Remove the 3.5 mm adapter from the setup.
4.
At this time, it is recommended that you [SAVE] the present setup to create a
State File (*.csa) on the network analyzer for RECALL, if needed for future use.
5.
Minimize the PNA-X application, FILE > Minimize Application, to access the
Test Set command application.
6.
Select the Windows Start menu and select the "Agilent Connection Expert."
7. When the Connection Expert application appears, close the Welcome screen (if it appears)
and look for the Test Set identified as L449XA. If L449XA is not showing, go to the top of
the screen and select Add Instrument.
- Enter GPIB address in the field for your instrument
- Check the two boxes for:
- Check this GPIB Address
- Auto ID this Instrument
- OK
Information related to the L449XA or U3020A will be displayed on the right side of the
window.
Figure
6
Connection Expert, Connected Instruments
U3020AD01
Operational Verification
Keysight U3020AD01 User's and Service Guide
12