Test Record Forms
Keysight RP7951A/RP7961A
RP7951A/RP7961A
Test Record
Report Number ______
Date__________
Test Description
Model
Min. Specs
Results
Max. Specs
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
0.43985
Vout - 0.08015
499.79
Vout - 0.230
_______
_______
_______
_______
0.56015
Vout + 0.08015
500.21
Vout + 0.230
CV Load Effect:
Both
- 0.030
_______
+ 0.030
CV Ripple and Noise
peak-to-peak:
rms:
Both
Both
N/A
N/A
_______
_______
0.500
0.100
Transient response @ 500 µs:
Both
- 1.25
_______
+ 1.25
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
Both
Both
Both
Both
- 0.012
Iout - 0.012
19.968
Iout - 0.032
_______
_______
_______
_______
0.012
Iout + 0.012
20.032
Iout + 0.032
CC Load Effect:
Both
- 0.009
_______
+ 0.009
Current Sink Tests
100% of current rating:
Both
- 20.032
_______
- 19.968
RP7951A/RP7961A UUT Settings
RP7952A/RP7962A Load Settings
Voltage Programming & Readback Min:
Voltage Programming & Readback High:
CV Load Effect, CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min:
Current Programming & Readback High:
CC Load Effect:
100% of current rating (Isink):
Voltage priority: 0.5 V, 10.25 A
Voltage priority: 500 V, 10.25 A
Voltage priority: 250 V, 20.5 A
Voltage priority: 250 V, 20.5 A
Current priority: 255 V, 0 A
Current priority: 255 V, 20 A
Current priority: 510 V, 10 A
Current priority: 255 V, - 20 A
not used
not used
Current priority: 255 V, 20 A
Current priority: 255 V, 10 A to 20 A, slew=max
Voltage priority: 250 V, 0 A
Voltage priority: 250 V, 20.5 A
Voltage priority: 500 V, 10.25 A
Voltage priority: 250 V, 20.5 A
Keysight RP7900 Series Operating and Service Guide
279
6 Verification and Calibration
Содержание RP7900 Series
Страница 1: ...Keysight Regenerative Power System RP7900 Series Operating and Service Guide ...
Страница 2: ......
Страница 51: ...Dimensions RP793xA and RP794xA Keysight RP7900 Series Operating and Service Guide 51 1 Quick Reference ...
Страница 60: ...Dimensions RP795xA and RP796xA 1 Quick Reference 60 Keysight RP7900 Series Operating and Service Guide ...
Страница 100: ......
Страница 244: ...Status Diagram 5 SCPI Programming Reference 244 Keysight RP7900 Series Operating and Service Guide ...
Страница 246: ...Trigger Diagram 5 SCPI Programming Reference 246 Keysight RP7900 Series Operating and Service Guide ...
Страница 262: ......
Страница 314: ......