Parameter
Typical Return
0 | 1 | 2
0, 1, or 2
10 to 100,000 for RP793xA and RP794xA all ranges
200 to 500,000 for RP795xA and RP796xA range 0 (Low)
17,000 to 500,000 for RP795xA and RP796xA range 1 (High1)
<pole frequency>
Sets the voltage bandwidth to 1, frequency 10 kHz: VOLT:BWID:LEV 1, 10,000
[SOURce:]VOLTage:BWIDth:RANGe 0 | 1 | 2
[SOURce:]VOLTage:BWIDth:RANGe?
Sets the voltage compensation. This lets you optimize output response time with capacitive loads.
These compensation modes only apply when the unit is operating in voltage priority mode. Refer to
for information about the effect that the compensation settings have on the
Programming Speed Characteristics.
0 (High speed/Small capacitive load) – provides the fastest programming speed and transient
response time.
1 (Medium speed/Medium capacitive load) – provides intermediate programming speed and transient
response time. setting is optimized for stability with a wide range of output capacitors.
2 (Slow speed/Large capacitive load) – best suited for DUTs with high capacitance/low ESR with the
trade-off of a slower programming speed and transient response.
Parameter
Typical Return
0 | 1 | 2, *RST 0
0, 1, or 2
Sets the voltage bandwidth to comp 1: VOLT:BWID:RANG 1
[SOURce:]VOLTage:LIMit[:POSitive][:IMMediate][:AMPLitude] <
value
>|MIN|MAX
[SOURce:]VOLTage:LIMit[:POSitive][:IMMediate][:AMPLitude]? [MIN|MAX]
Sets the voltage limit when in current priority mode. Units are in volts.
Parameter
Typical Return
0.1% to 102% of rating, *RST 1% of rating
<voltage limit>
Sets the voltage limit to 20 V: VOLT:LIM 20
[SOURce:]VOLTage:LIMit:LOW <
value
>|MIN|MAX
[SOURce:]VOLTage:LIMit:LOW? [MIN|MAX]
Sets the low voltage limit when in current priority mode. This prevents the voltage from dropping
below the low voltage limit when discharging a battery. When the voltage drops to the specified low
limit value, the output transitions from current priority to negative voltage limit and the discharging
stops. This sets the LIM- bit in the Questionable Status Register.
5 SCPI Programming Reference
236
Keysight RP7900 Series Operating and Service Guide
Содержание RP7900 Series
Страница 1: ...Keysight Regenerative Power System RP7900 Series Operating and Service Guide ...
Страница 2: ......
Страница 51: ...Dimensions RP793xA and RP794xA Keysight RP7900 Series Operating and Service Guide 51 1 Quick Reference ...
Страница 60: ...Dimensions RP795xA and RP796xA 1 Quick Reference 60 Keysight RP7900 Series Operating and Service Guide ...
Страница 100: ......
Страница 244: ...Status Diagram 5 SCPI Programming Reference 244 Keysight RP7900 Series Operating and Service Guide ...
Страница 246: ...Trigger Diagram 5 SCPI Programming Reference 246 Keysight RP7900 Series Operating and Service Guide ...
Страница 262: ......
Страница 314: ......