Parameter
Typical Return
0|OFF|1|ON
0 or 1
Enables the SDS unit: SYST:SDS:ENAB ON
SYSTem:SDS:STATus?
Returns the protection status of the SDS unit. The integer bits indicate the reason for the SDS
protection.
Bit
Description
0
A power supply-to-SDS communication failure
1
AC contact fault - the AC contact input is not active
2
An Internal SDS hardware failure
These faults can only be cleared by turning off the power supply, correcting the problem and rebooting
the power supply.
Parameter
Typical Return
(none)
<integer>
Return the status of the SDS unit: SYST:SDS:STAT?
SYSTem:SECurity:IMMediate
Clears all user memory and reboots the instrument. This command is typically used to prepare the
instrument for removal from a secure area. It sanitizes all user data. It writes all zeros to flash memory
and then performs a chip erase as per manufacturer's data sheet. Identification data (instrument
firmware, model number, serial number, MAC address and calibration data) is not erased. After the
data is cleared, the instrument is rebooted.
This procedure is not recommended for use in routine applications because of the possibility of
unintended loss of data.
Parameter
Typical Return
(none)
(none)
Sanitizes the instrument: SYST:SEC:IMM
SYSTem:SET <block data>
SYSTem:SET?
Get and set the instrument state. The query form of the command returns a definite length block that
contains the instrument state. This block data can be sent back to the instrument to restore it to the
state in the block.
5 SCPI Programming Reference
230
Keysight RP7900 Series Operating and Service Guide
Содержание RP7900 Series
Страница 1: ...Keysight Regenerative Power System RP7900 Series Operating and Service Guide ...
Страница 2: ......
Страница 51: ...Dimensions RP793xA and RP794xA Keysight RP7900 Series Operating and Service Guide 51 1 Quick Reference ...
Страница 60: ...Dimensions RP795xA and RP796xA 1 Quick Reference 60 Keysight RP7900 Series Operating and Service Guide ...
Страница 100: ......
Страница 244: ...Status Diagram 5 SCPI Programming Reference 244 Keysight RP7900 Series Operating and Service Guide ...
Страница 246: ...Trigger Diagram 5 SCPI Programming Reference 246 Keysight RP7900 Series Operating and Service Guide ...
Страница 262: ......
Страница 314: ......