Parameter
Typical Return
(none)
50 or 60
Query the power line frequency: SYST:LFR?
SYSTem:LFRequency:MODE AUTO|MAN50|MAN60
SYSTem:LFRequency:MODE?
Specifies automatic or manual line frequency detection.
AUTO specifies automatic detection.
MAN50 specifies a setting of 50 Hz.
MAN60 specifies a setting of 60 Hz.
This parameter is saved in non-volatile memory.
Parameter
Typical Return
AUTO|MAN50|MAN60
AUTO, MAN50, or MAN60
Sets the line frequency mode to 60 Hz: SYST:LFR:MODE MAN60
SYSTem:PASSword:FPANel:RESet
Resets the front panel lockout password to zero. This command does not reset the calibration
password.
Parameter
Typical Return
(none)
<+0,"No error">
Resets the front panel password: SYST:PASS:FPAN:RES
SYSTem:REBoot
Reboots the instrument to its power-on state.
Parameter
Typical Return
(none)
(none)
Reboots the instrument: SYST:REB
SYSTem:SDS:CONNect
Connects the power supply to the Keysight SD1000A SDS unit. This command is used when the SDS
connect mode is set to MANual.
Parameter
Typical Return
(none)
(none)
Connect the SDS unit: SYST:SDS:CONN
5 SCPI Programming Reference
228
Keysight RP7900 Series Operating and Service Guide
Содержание RP7900 Series
Страница 1: ...Keysight Regenerative Power System RP7900 Series Operating and Service Guide ...
Страница 2: ......
Страница 51: ...Dimensions RP793xA and RP794xA Keysight RP7900 Series Operating and Service Guide 51 1 Quick Reference ...
Страница 60: ...Dimensions RP795xA and RP796xA 1 Quick Reference 60 Keysight RP7900 Series Operating and Service Guide ...
Страница 100: ......
Страница 244: ...Status Diagram 5 SCPI Programming Reference 244 Keysight RP7900 Series Operating and Service Guide ...
Страница 246: ...Trigger Diagram 5 SCPI Programming Reference 246 Keysight RP7900 Series Operating and Service Guide ...
Страница 262: ......
Страница 314: ......