Trigger Source
Description
Bus
Selects GPIB device trigger, *TRG, or <GET> (Group Execute Trigger).
Current
Selects an output current level.
External
Selects ANY pin that has been configured as a Trigger Input on the digital control
port.
Pin<1-7>
Selects a specific pin<n> that is configured as a Trigger Input on the digital con-
trol port.
Transient
Selects the unit’s transient system. You must also set up the transient system to
generate a trigger out signal. See
.
Voltage
Selects an output voltage level.
Use the following commands to select a trigger source:
Front Panel Menu Reference
SCPI Command
Not available
To select Bus triggers:
To select digital pin 5 as the trigger:
To select a voltage or current level:
TRIG:ACQ:SOUR VOLT
TRIG:ACQ:SOUR CURR
To select an output transient as trigger:
Initiate the acquisition system
When the unit is turned on, the trigger system is in the idle state. In this state, the trigger system is
disabled, ignoring all triggers. The INITiate commands enable the trigger system to receive triggers.
Front Panel Menu Reference
SCPI Command
Not available
To initiate the measurement trigger system:
It takes a few milliseconds for the instrument to be ready to receive a trigger signal after receiving the
INITiate:ACQuire command. If a trigger occurs before the trigger system is ready for it, the trigger will
be ignored. You can test the WTG_meas bit in the operation status register to know when the
instrument is ready to receive a trigger after being initiated.
Front Panel Menu Reference
SCPI Command
Select
Measure\Control
.
The Trig state field indicates "Initiated".
To query the WTG_meas bit (bit 3):
Keysight RP7900 Series Operating and Service Guide
151
4 Using the Regenerative Power System
Содержание RP7900 Series
Страница 1: ...Keysight Regenerative Power System RP7900 Series Operating and Service Guide ...
Страница 2: ......
Страница 51: ...Dimensions RP793xA and RP794xA Keysight RP7900 Series Operating and Service Guide 51 1 Quick Reference ...
Страница 60: ...Dimensions RP795xA and RP796xA 1 Quick Reference 60 Keysight RP7900 Series Operating and Service Guide ...
Страница 100: ......
Страница 244: ...Status Diagram 5 SCPI Programming Reference 244 Keysight RP7900 Series Operating and Service Guide ...
Страница 246: ...Trigger Diagram 5 SCPI Programming Reference 246 Keysight RP7900 Series Operating and Service Guide ...
Страница 262: ......
Страница 314: ......