73
Functions and Measurements
Reflection Measurement
connect the Open component of the T-combo calibrator to the
RF OUT connector of the analyzer. Press
ENTER
to start the open
calibration.
2 Follow the instruction to connect the short component of the
T-combo calibrator to the RF OUT connector of the tester. Press
ENTER
to start the short calibration.
3 Connect the Load component of the T-combo calibrator to the
RF OUT connector.
Press
ENTER
to start the load calibration.
A
Calibrated
indicator is shown on the upper left corner of the
screen when the calibration is completed. Then, the analyzer is
ready for your further cable measurement.
Measuring Cable Reflection
Press
Mode > Reflection Measurement > Reflection Measurement
to
select the Reflection Measurement Mode. It allows you to set the Y
axis of the graticule as return loss, VSWR or reflection coefficient
value.
Measuring Distance to Fault
For this measurement, an internally generated signal emits from
the RF OUT port of the tester and is transmitted to the cable and
antenna system (DUT). The tester receives and processes the
signal power reflected from faults and imperfections. In performing
a DTF measurement, the tester uses frequency domain
reflectometry. The transmitted and reflected signals contain
information about the distance to the faults. This information is
used to determine the physical distance to the faults. The tester
displays the physical distance to the probable faults, corrected for
cable loss and the velocity propagation factor of the cable.
Содержание N9321C
Страница 36: ...Getting Started Helpful Tips 28 ...
Страница 110: ...Functions and Measurements Files Operation 102 ...
Страница 180: ...Key Reference Trace Det 172 ...
Страница 212: ...Menu Maps Name Editor Menu 204 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Страница 214: ...Menu Maps SPAN Menu 206 SPAN Menu Span Span Full Span Zero Span Last Span ...
Страница 216: ...Menu Maps Trace Det Menu 208 ...
Страница 219: ......