49
Functions and Measurements
Improving Frequency Resolution and Accuracy
Improving Frequency Resolution and Accuracy
This section provides information on using the frequency counter
function to improve frequency resolution and accuracy.
1
Press
Preset
(With
Preset Type
of
DFT
)
2
Connect a cable from the front panel
CAL OUT
to
RF IN
;
Press
System > Service > Calibration > CAL OUT ON
to toggle on and
enable the 50 MHz amplitude reference signal.
3
Press
Frequency > Auto Tune
.
The analyzer will detect the signal peak and locate it to the
center
of
the
display (Refer
to
4
Turn the frequency counter on:
Press
Marker > Function > Counter
.
5
Move the marker, with the front-panel knob, half-way down the
skirt of the signal response.
Figure 3-14Using Frequency Counter
NOTE
The frequency and amplitude of the marker appears in the active
function area (this is not the counted result). The counted result
appears in the upper-right corner of the display to the right-side of
C1.
Counted
result