Instrument Messages
Device-Specific Errors
Device-Specific Errors
102
Set 50 MHz signal < 0 dBm, do alignment again
The 50 MHz energy is too high at the RF port for alignments to run.
Reduce input power to less than 0 dBm and run alignments again.
104
1st LO Unlock
The first LO on has lost phase lock. Report this error to the nearest
Keysight Technologies sales or service office.
105
2nd LO Unlock
The second LO has lost phase lock. Report this error to the nearest
Keysight Technologies sales or service office.
201
Option not installed
The desired operation cannot be performed because a required
option is not installed.
204
TG Frequency Limit
The tracking generator has reached the limit of its allowable
frequency range.
211
RBW limited to 1 kHz when Span > 1 MHz
In spans greater than 1 MHz, narrow (digital) resolution
bandwidths, below 1 kHz, are not available.
213
Span limited to 1 MHz when RBW < 1 kHz
In narrow (digital) resolution bandwidths, below 1 kHz, spans
greater than 1 MHz are not available.
214
TG start freq is less than 9 kHz
Tracking generator uncalibrated below 9 kHz.
215
TG start freq is less than 1/2 res bw
Tracking generator uncalibrated at start frequencies below 1/2 the
current resolution bandwidth.
Содержание N9321C
Страница 36: ...Getting Started Helpful Tips 28 ...
Страница 110: ...Functions and Measurements Files Operation 102 ...
Страница 180: ...Key Reference Trace Det 172 ...
Страница 212: ...Menu Maps Name Editor Menu 204 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Страница 214: ...Menu Maps SPAN Menu 206 SPAN Menu Span Span Full Span Zero Span Last Span ...
Страница 216: ...Menu Maps Trace Det Menu 208 ...
Страница 219: ......