Key Reference
MODE
134
MODE
Selects the measurement mode of your analyzer.
Spectrum Analyzer
mode is the default mode and is for general
purpose measurement use.
Tracking Generator
Accesses the tracking generator functions. (Option TG3)
Key access:
Mode
Amplitude
Toggles the tracking generator output between on and off. The
power level can then be adjusted using the numeric keypad, arrow
keys or knob. Pressing any digit, 0 through 9, on the numeric
keypad brings up the selected terminator menu. The output level
range is –30 dBm to 0 dBm. When Amplitude (Off) is selected, the
output attenuation is set to the maximum attenuation.
Key access:
Mode > Tracking Generator
Amptd Step
Sets the step size of the power level range of the tracking
generator. The default setting is one vertical scale division when in
a logarithmic amplitude scale.
Key access:
Mode > Tracking Generator
CAUTION
Power-level sensitive devices connected to the tracking generator
output may be accidentally damaged. This is because the actual source
amplitude will be greater than the amplitude indicated on the analyzer,
when the power sweep function is being used. The source amplitude
value is used as the starting amplitude for a power sweep. The ending
amplitude for a power sweep is higher then the source amplitude value
and it may actually be set to values beyond specified output levels.
Therefore, exercise caution when connecting a power-level sensitive
device to the tracking generator output.
Содержание N9321C
Страница 36: ...Getting Started Helpful Tips 28 ...
Страница 110: ...Functions and Measurements Files Operation 102 ...
Страница 180: ...Key Reference Trace Det 172 ...
Страница 212: ...Menu Maps Name Editor Menu 204 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Страница 214: ...Menu Maps SPAN Menu 206 SPAN Menu Span Span Full Span Zero Span Last Span ...
Страница 216: ...Menu Maps Trace Det Menu 208 ...
Страница 219: ......