N9030B PXA Signal Analyzer Service Guide 99
Instrument Messages
Event Messages
–221 Settings Conflict Errors
This is one of the errors in the standard SCPI error range of –200 to –299. See
the table
“–200 to –299, Execution Errors” on page 110
The <subtext> part of a Settings Conflict error should be worded so that the
text is: “function1” is not whatever/with/while/when “function2”. This makes
them easier to find - alphabetically, to avoid duplicates.
The entire message displays in the error history as “–221, Settings conflict;
<subtext>”
For example, –221.0076 displays as:
–221, Settings conflict; Invalid trace number
Err#
Message
Verbose/Correction Information
–221
De-emphasis only available
in FM
The de-emphasis function is only available if FM demod is
selected.
–221
Function not available in
Zero Span
The function you are trying to access is not available in zero
span.
–221
Setting conflict; <trigger
source> trigger is not
available while input is
<input port>
The trigger source (Video, RF Burst, I/Q Mag, etc.) is not
available with the current input port (RF, IQ, etc.)
–221
Setting conflict;
Differential setting
determined by probe type
A probe is connected that has a built in Differential setting.
The setting cannot be changed manually.
–221
Setting conflict; Input Z
unavailable when probe
sensed
A probe is connected and the Input Z is set based on the
probe type. It cannot be changed manually.
–221
Settings conflict;
A legal command was received but it could not be executed
due to the current device state.
–221
Settings conflict; *.CSV
file format is not available
in this measurement.
You cannot load or save base instrument traces, as this is not
supported by the Log Plot measurement.
–221
Settings conflict; <Q Param>
cannot be changed when Q
same as I
When the “Q Same as I” parameter is set to Yes, the I
parameter value is copied to <Q Param> and the <Q Param>
value cannot be changed. Set Q Same as I to No to enable
explicit control of the <Q Param> value.
–221
Settings conflict; A Valid
User Cal is required.
Optimize aborted
Optimize Preselector can only be performed if a valid user
cal exists and is applied to current results. Perform a user
cal first or apply existing cal.
–221
Settings conflict; Auto Tune
not available in Tracking
Source mode
The Auto Tune feature cannot be used when you are using a
Tracking Source.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...