Performance Tests
2
Keysight N8262A Service Guide
47
Test method
1
Obtain the S11 parameter of the Calibration System.
2
Connect the Measurement System to the Calibration System and obtain its
S21 (load) and S21 (open) parameters.
3
Using only the Measurement System, terminated with the OPEN connector
from the 85032B calibration kit, measure the 1 mW power reference level of
the DUT.
4
Remove the OPEN connector from the Measurement System, terminate it with
the 50 R load from the 85032B calibration kit, and repeat the 1 mW power
reference level measurement.
5
Using all of these values, the N7832A software will calculate the VSWR of the
power reference output.
NOTE
– This test cannot be performed manually, due to the complexity of the
equipment calibration procedure and the complexity of the measurement
algorithm.
– No adjustment is available for this test if it fails (see
Содержание N8262A
Страница 1: ...Keysight N8262A P Series Modular Power Meter Service Guide ...
Страница 18: ...18 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 40: ...1 Specifications 40 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 58: ...2 Performance Tests 58 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 62: ...3 Adjustments 62 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 70: ...4 Theory of Operation 70 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...