General Power Meter Functions
2
Keysight N8262A User’s Guide
83
Setting limits
The power meter can be configured to verify the current measurement in any
measurement line against predefined upper and/or lower limit values. The range
of values that can be set for the upper and lower limits and the default values
depends on the measurement units in the currently selected measurement line
).
Procedure
Set the limits as follows:
1
Click
,
to display the
Measurement Setup
menus.
2
Use the
and
softkeys to highlight the
Limits:
setting field.
3
Click
to check the
Limits:
setting field.
4
Use the
softkey to highlight the
Minimum Limits:
value field.
5
Click
to display the
Minimum Limit
pop-up.
Table 2-4
Range of values for window limits
Window units
Maximum
Minimum
Default maximum Default minimum
dB
+200 dB
–180 dB
60 dB
–120 dB
dBm
+230 dBm
–150 dBm
90 dBm
–90 dBm
%
999.9 X%
100.0 a%
100.0 M%
100.0 p%
W
100.000 XW
1.000 aW
1.000 MW
1.000 pW
NOTE
Ensure you have selected the channel you wish to set up.
Meas Select
Содержание N8262A P Series
Страница 1: ...Keysight N8262A P Series Modular Power Meter User s Guide ...
Страница 18: ...18 Keysight N8262A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight N8262A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 106: ...2 General Power Meter Functions 106 Keysight N8262A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 146: ...5 Characteristics and Specifications 146 Keysight N8262A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...