2
General Power Meter Functions
64
Keysight N8262A User’s Guide
Setting the Video Averaging
Video averaging uses a digital filter to average repetitions of a triggered signal.
The number of acquisitions averaged can range from 1 to 256, in multiples of 2
n
.
With video averaging the average of a number of acquisitions is calculated to
smooth the displayed trace and reduce apparent noise. The measurement
requires a continuously repeating signal. Increasing the value of this filter reduces
noise but increases the time required to make the measurement.
Procedure
Set the video average as follows:
1
Click
. Click
to select the required channel.
2
Use the
and
softkeys to select the
Video Avg
setting field.
3
Click
to check the video averaging to on or off.
4
Use the
softkey to select the
Video Avg:
value field.
5
Click
to display the
Video Avg Count
pop-up.
NOTE
This feature is only available when a P-Series Power Sensor is connected.