![Keysight Technologies N778 C Series Скачать руководство пользователя страница 89](http://html.mh-extra.com/html/keysight-technologies/n778-c-series/n778-c-series_programming-manual_1962052089.webp)
Keysight N778xC Series Polarization Instruments Programming Guide
89
Polarization Commands
5
Polarization Synthesizer / Polarization Component Analyzer Commands
Command:
:STABilizer:STABilize
Syntax:
:STABilizer:STABilize<wsp> <value>
Description:
Sets the Stabilizer Operating Mode.
NOTE - The stabilizer function configures and controls both, the Scrambler and the Polarimeter including intern and extern
Trigger Configuration. After Stopping the Stabilizer Mode use System:Preset or *RST to activate the normal Scrambler
and Polarimeter Mode
NOTE - Use only gain between 0 and 7 for Stabilizer Mode because 8 and 9 have a small Bandwith e.g. POL:GAIN 6
Auto Gain is disabled when using Logging / Sweep and Stabilizer Mode For best results check the Leveling of the
Polarimeter. Look at the POL:SWE:LPR? command.
Maximum Speed is at Gain 0 to 5
Even Gains have Factors of 10 and odd Gains Factors of 8. Choose what fits better.
NOTE - If the stabilizer function is active then most other Commands to Scrambler and Polarimeter including Zeroing are
blocked.
NOTE - An "On Startup Stabilize Mode" where stabilizing is automatically on after reboot can be enabled by saving the
Configuration in one of the 10 Setting Positions e.g. with following commands:
:stab:sop 1,0,0
:stab:stab 1
:conf:meas:sett:save 1
Parameters:
0 = Stabilizer Off (Default)
1 = Stabilizer On
Response:
None
Example:
:STAB:STAB 1
Affects
N7785C, N7786C, N7788C
Command:
:STABilizer:STABilize?
Syntax:
:STABilizer:STABilize?
Description:
Gets the Stabilizer Operating Mode.
Parameters:
None
Response:
0 = Stabilizer Off (Default)
1 = Stabilizer On
Example:
:STAB:STAB?
→
1
Affects
N7785C, N7786C, N7788C