0 Fast (high speed/small capacitive load) – provides the fastest programming speed and transient
response time. Best suited for resistive DUTs with low capacitance.
1 Medium (medium speed/medium capacitive load) – provides intermediate programming speed and
transient response time. Best suited for DUTs with higher capacitance with some trade-off in
programming speed and transient response.
2 Slow (slow speed/large capacitive load) – is best suited for DUTs with high capacitance/low ESR
with the trade-off of slower programming speed and transient response.
The following table summarizes the effect that the compensation settings have on the CV
Programming Speed Characteristics for all models.
Setting
Step Conditions
Rise/fall time
10%-90% of step
Settling time
to 0.1% of step
Comp 0 Fast
no load, 0.1%-100% of rating
10 ms
50 ms
Comp 1 Medium
no load, 0.1%-100% of rating
20 ms
100 ms
Comp 2 Slow
no load, 0.1%-100% of rating
40 ms
200 ms
Front Panel Menu Reference
SCPI Command
Select
Output\Advanced\Bandwidth\Voltage
.
Select either 0, 1, or 2.
To set the voltage compensation bandwidth:
Current compensation - only applies to models N69143A, N69145A, and N69146A
Current compensation for models N6914xA lets you optimize output response time with inductive
loads. This compensation is suited to handle a wide, range of total lead and DUT inductance with a
relatively fast programming response (see
graph). To reduce the impact of
load lead inductance, paralleling additional sets of load leads to lower the net inductance can help,
especially if load lead lengths are long.
Comp 0 Best suited for longer load leads (higher inductance) with relatively fast programming
response.
Comp 1 Optimizes programming speed for DUTs with limited lead inductance. This setting may require
paralleling additional sets of leads to achieve the desired overall lead length.
Setting
Step Conditions
Rise/fall time
10%-90% of step
Settling time
to 0.1% of step
Comp 0
0 -100% of current step at Vout
2.5 ms
5 ms
Comp 1
0 -100% of current step at Vout
5 ms
10 ms
Front Panel Menu Reference
SCPI Command
Select
Output\Advanced\Bandwidth\Current
.
Select either 0, or 1.
To set the current compensation bandwidth:
Keysight N69100 Series Operating and Service Guide
99
4 Using the DC Power System
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......