![Keysight Technologies N69100 Series Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/keysight-technologies/n69100-series/n69100-series_operating-and-service-manual_1962218036.webp)
Common Characteristic
All Models
Acoustic statement
(European Machinery Directive)
Acoustic noise emission
LpA 80.92 dB at Operator position:1
LpA 75.57 dB at Bystander position
LpA 62.04 at Idle fan speed
Normal operation mode per ISO 7779
Environmental
Operating environment:
Temperature range:
Relative humidity:
Altitude:
Storage temperature:
Indoor use, installation category III, pollution degree 2
0°C to 55°C (Maximum continuous current available is derated at 1% of rating per degree C from
40°C to 55°C)
95% or less (non-condensing)
Up to 2000 meters
-30°C to 70°C
Regulatory Compliance
EMC:
Safety:
Complies with European EMC Directive for test and measurement products
Complies with Australian standard and carries C-Tick mark
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada
Complies with European Low Voltage Directive and carries the CE mark.
Conforms to US and Canadian safety regulations.
Declarations of Conformity for this product may be downloaded from the Web. Go to
http://www.keysight.com/go/conformity
and click on “Declarations of Conformity.”
AC Input - N6914xA units
Connections:
Phase and range:
Frequency:
Input VA:
Input current per phase
Efficiency at full power:
Power Factor:
L1, L2, L3, PE (no neutral connection required)
3 phase; 400-480 VAC nominal ±10%
50/60 Hz
11.5 kVA
17.3 A
85%
0.99 @ nominal input and rated power
AC Input - N6917xA units
Connections:
Phase and range:
Frequency:
Input VA:
Input current per phase
Efficiency at full power:
Power Factor:
L1, L2, L3, N, PE (this requires a neutral connection)
3 phase; 380-480 VAC nominal ±10%
50/60 Hz
23 kVA
36 A
90%
0.99 @ nominal input and rated power
Output Terminal Isolation
for 20 V units:
for 80 V & 160 V units:
for 1 kV units:
for 2 kV units:
No output terminal may be more than ±60 VDC from any other terminal or chassis ground.
No output terminal may be more than ±240 VDC from any other terminal or chassis ground.
No output terminal may be more than ±1000 VDC from any other terminal or chassis ground.
No output terminal may be more than ±2000 VDC from any other terminal or chassis ground.
Typical Weight
for N6914xA units:
for N6917xA units:
70 lbs. (31.8 kg)
82 lbs. (37.3 kg)
Dimensions
3U high; full rack width (see
for details)
1 A sound pressure level of 80 dBA above a reference sound pressure of 20 μPa is at present regarded by many authorities as the threshold at
which a hazard may be caused. Special means, such as the use of protective earpieces, can make a higher level non-hazardous to an operator.
1 Quick Reference
36
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......