TRIGger:ARB:SOURce <
source
>
TRIGger:ARB:SOURce?
TRIGger:TRANsient:SOURce <
source
>
TRIGger:TRANsient:SOURce?
TRIG:ARB:SOURce - Selects the trigger source for arbitrary waveforms:
TRIG:TRANsient:SOURce - Selects the trigger source for the transient system:
BUS
Selects a remote interface trigger command.
EXTernal
Selects ALL digital port pins that have been configured as trigger sources.
IMMediate
Triggers the transient as soon as it is INITiated.
PIN<1-7>
Selects a digital port pin configured as a trigger input.
Parameter
Typical Return
BUS | EXTernal | IMMediate | PIN<1-7> *RST BUS
BUS, EXT, IMM, PIN<n>
Select digital port pin 1 as the Arb trigger source: TRIG:ARB:SOUR PIN1
Select digital port pin 1 as the transient trigger source: TRIG:TRAN:SOUR PIN1
5 SCPI Programming Reference
202
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......