[SOURce:]LIST:TOUTput:BOSTep[:DATA] 0|OFF|1|ON {,0|OFF|1|ON}
[SOURce:]LIST:TOUTput:BOSTep[:DATA]?
[SOURce:]LIST:TOUTput:EOSTep[:DATA] 0|OFF|1|ON {,0|OFF|1|ON}
[SOURce:]LIST:TOUTput:EOSTep[:DATA]?
Specifies which list steps generate a trigger signal at the beginning of step (BOSTep) or end of step
(EOSTep). A trigger is only generated when the state is set to ON. The trigger signal can be used as a
trigger source for measurements and transients of other units, and for digital port pins configured as
trigger outputs.
Parameter
Typical Return
0|OFF|1|ON
0 or 1
To generate triggers at the beginning of the second step of a 3-step list:
LIST:TOUT:BOST OFF,ON,OFF
5 SCPI Programming Reference
180
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series