[SOURce:]VOLTage:LIMit[:POSitive][:IMMediate][:AMPLitude] <
value
>|MIN|MAX
[SOURce:]VOLTage:LIMit[:POSitive][:IMMediate][:AMPLitude]? [MIN|MAX]
Sets the voltage limit when in current priority mode. Units are in volts.
Parameter
Typical Return
0.1% to 102% of rating, *RST 1% of rating
<voltage limit>
Sets the voltage limit to 20 V: VOLT:LIM 20
[SOURce:]VOLTage:MODE FIXed|STEP|LIST|ARB
[SOURce:]VOLTage:MODE?
Sets the transient mode. This determines what happens to the output voltage when the transient
system is initiated and triggered.
FIXed keeps the output voltage at its immediate value.
STEP steps the output to the triggered level when a trigger occurs.
LIST causes the output to follow the list values when a trigger occurs.
ARB causes the output to follow the arbitrary waveform values when a trigger occurs.
Parameter
Typical Return
FIXed|STEP|LIST|ARB, *RST FIXed
FIX, STEP, LIST, or ARB
Sets the voltage mode to Step: VOLT:MODE STEP
[SOURce:]VOLTage:PROTection[:LEVel] <
value
>|MIN|MAX
[SOURce:]VOLTage:PROTection[:LEVel]? [MIN|MAX]
Sets the over-voltage protection level. Units are in volts. If the output voltage exceeds the OVP level,
the output is disabled and the Questionable Condition status register OV bit is set.
Parameter
Typical Return
0 to 120% of rating, *RST 120% of rating
<protect level>
Sets the over-voltage protection to 24 V: VOLT:PROT 24
l
An over-voltage condition can be cleared with the OUTput:PROTection:CLEar command after the
cause of the condition has been removed.
[SOURce:]VOLTage:RESistance[:LEVel][:IMMediate][:AMPLitude] <
value
>|MIN|MAX
[SOURce:]VOLTage:RESistance[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX]
Sets the output resistance level in voltage in voltage priority mode. Units are in ohms. Refer to
for the model-specific resistance programming ranges.
Keysight N6900/N7900 Series Operating and Service Guide
282
6 SCPI Programming Reference
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...