VOLTage Subsystem
Voltage commands program the output voltage of the instrument.
[SOURce:]
VOLTage
[:LEVel]
[:IMMediate]
[:AMPLitude] <
value
>
Sets the output voltage when in voltage priority mode.
:TRIGgered
[:AMPLitude] <
value
>
Sets the triggered output voltage.
:BWIDth LOW|HIGH1
Sets the voltage bandwidth.(for backward compatibility)
:LEVel 0|1, <
value
>
Sets the programming pole frequency.
:RANGe 0|1
Sets the voltage bandwidth.
:LIMit
[:POSitive]
[:IMMediate]
[:AMPLitude] <
value
>
Sets the voltage limit when in current priority mode.
:MODE FIXed|STEP|LIST|ARB
Sets the transient mode.
:PROTection
[:LEVel] <
value
>
Sets the over-voltage protection level.
:RESistance
[:LEVel]
[:IMMediate]
[:AMPLitude] <
value
>
Sets the output resistance when in voltage priority mode.
:STATe 0|OFF|1|ON
Enables or disables resistance programming in voltage pri-
ority mode.
:SLEW
[:IMMediate] <
value
>|INFinity
Sets the voltage slew rate.
:MAXimum 0|OFF|1|ON
Enables or disables the maximum slew rate override.
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <
value
>|MIN|MAX
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX]
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <
value
>|MIN|MAX
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX]
Sets the immediate or triggered voltage level when the output is operating in voltage priority mode.
The triggered level is the value that is transferred to the output when an output Step is triggered.
Units are in volts. The maximum value depends on the voltage rating of the unit.
Parameter
Typical Return
0.1% to 102% of rating, *RST 0.1% of rating
<voltage level>
Sets the output voltage to 20 V: VOLT 20
Keysight N6900/N7900 Series Operating and Service Guide
280
6 SCPI Programming Reference
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...