Parameter
Typical Return
AUTO|MAN50|MAN60
AUTO, MAN50, or MAN60
Sets the line frequency mode to 60 Hz: SYST:LFR:MODE MAN60
SYSTem:PASSword:FPANel:RESet
Resets the front panel lockout password to zero. This command does not reset the calibration
password.
Parameter
Typical Return
(none)
(none)
Resets the front panel password: SYST:PASS:FPAN:RES
SYSTem:REBoot
Reboots the instrument to its power-on state.
Parameter
Typical Return
(none)
(none)
Reboots the instrument: SYST:REB
SYSTem:SECurity:IMMediate
Clears all user memory and reboots the instrument. This command is typically used to prepare the
instrument for removal from a secure area. It sanitizes all user data. It writes all zeros to flash memory
and then performs a chip erase as per manufacturer's data sheet. Identification data (instrument
firmware, model number, serial number, MAC address and calibration data) is not erased. After the
data is cleared, the instrument is rebooted.
This procedure is not recommended for use in routine applications because of the possibility of
unintended loss of data.
Parameter
Typical Return
(none)
(none)
Sanitizes the instrument: SYST:SEC:IMM
l
On units with serial numbers prior to MY59100101, this command does not sanitize the BBR drive
(Keysight N7908A). The BBR drive must be removed and destroyed to sanitize the BBR data.
Keysight N6900/N7900 Series Operating and Service Guide
274
6 SCPI Programming Reference
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...