![Keysight Technologies N6900 Series Скачать руководство пользователя страница 238](http://html1.mh-extra.com/html/keysight-technologies/n6900-series/n6900-series_operating-and-service-manual_1962217238.webp)
Parameter
Typical Return
(none)
0 (pass) or +1 (failed)
Perform self-test: *TST?
l
The power-on self-test is the same self-test performed by *TST.
l
*TST? also forces an *RST command.
*WAI
Pauses additional command processing until all pending operations are complete. See
for more
information.
Parameter
Typical Return
(none)
(none)
Wait until all pending operations complete. *WAI
l
*WAI can only be aborted by sending the instrument a Device Clear command.