SENSe:VOLTage[:DC]:RANGe[:UPPer] <
value
>, (@<chanlist>)
SENSe:VOLTage[:DC]:RANGe[:UPPer]? [MIN|MAX,] (@<chanlist>)
Selects a DC voltage measurement range. The value that you enter must be the highest value that
you expect to measure. Units are in volts. The instrument selects the range with the best resolution
for the value entered.
Parameter
Typical Return
0 to maximum|MIN|MAX, *RST MAX
<max voltage>
Selects the 2 V range: SENS:VOLT:RANG 2, (@1)
l
When queried, the returned value is the maximum DC voltage that can be measured on the range
that is presently set.
SENSe:VOLTage[:DC]:RANGe:AUTO 0|OFF|1|ON, (@<chanlist>)
SENSe:VOLTage[:DC]:RANGe:AUTO? (@<chanlist>)
Enables/disables seamless measurement autoranging.
Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
Enable voltage measurement autoranging: SENS:VOLT:RANG:AUTO ON, (@1)
l
Autoranging does not include the 10 μA range, which must be selected individually.
SENSe:WINDow[:TYPE] HANNing|RECTangular, (@<chanlist>)
SENSe:WINDow[:TYPE]? (@<chanlist>)
Selects the measurement window. This sets a signal conditioning function used in scalar DC
measurement calculations. Neither window function alters the instantaneous voltage or current data
returned in the measurement array.
Hanning - window is a "raised cosine" function. It is a signal conditioning function that reduces errors in
DC measurement calculations in the presence of periodic signals such as AC line ripple. This window
only works up to 4883 measurement points. The instrument will revert to a rectangular window when
the points exceed 4883.
Rectangular - window returns measurement calculations with no signal conditioning.
Parameter
Typical Return
HANNing|RECTangular, *RST RECTangular
RECT or HANN
Specifies a Hanning window function: SENS:WIND HANN, (@1)
7 SCPI Programming Reference
326
Keysight N6705C Operating and Service Guide
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......