Parameter
Typical Return
0|OFF|1|ON, *RST current OFF; voltage ON
0 or 1
Enables internal datalog measurements:
SENS:DLOG:FUNC:CURR ON, (@1)
SENS:DLOG:FUNC:VOLT ON, (@1)
SENSe:DLOG:FUNCtion:MINMax 0|OFF|1|ON
SENSe:DLOG:FUNCtion:MINMax?
Enables or disables logging of the minimum and maximum current or voltage values. Applies to all
channels.
Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
Enables MIN/MAX datalog values: SENS:DLOG:FUNC:MINM ON
l
Although data is logged to the data file at the logging interval, it is sampled at a rate of 20.48
microseconds. When the data is logged, the minimum, maximum, and average values are com-
puted from all of the samples taken during the logging interval. With MINMax disabled, only the
average values are entered into the datalog file. With MINMax enabled, all three values are
entered into the datalog file - thus tripling its size.
SENSe:DLOG:MARKer1:POINt <value>
SENSe:DLOG:MARKer1:POINt?
SENSe:DLOG:MARKer2:POINt <value>
SENSe:DLOG:MARKer2:POINt?
Sets the position of the data logger markers in seconds relative to the trigger position. A negative time
gives positions before the trigger. The marker positions are adjusted to the closest actual sample
point. Querying the marker positions after a FETCh:DLOG:CURRent? or FETCh:DLOG:VOLTage? query
will return the exact time point that corresponds to the returned measurement data.
Parameter
Typical Return
- 359,996,400 to +359,996,400
*RST 7.5 for marker 1; 27.5 for marker 2
<marker point>
Sets the marker 1 position in seconds: SENS:DLOG:MARK1:POIN 50
Sets the marker 2 position in seconds: SENS:DLOG:MARK2:POIN 150
SENSe:DLOG:OFFSet <value>
SENSe:DLOG:OFFSet?
Specifies the datalog trigger offset as a percent of the total datalog duration. This lets you specify the
percent of pre-trigger data that will be logged to the datalog file.
7 SCPI Programming Reference
318
Keysight N6705C Operating and Service Guide
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......