[SOURce:]CURRent:PROTection:DELay:STARt SCHange|CCTRans, (@<chanlist>)
[SOURce:]CURRent:PROTection:DELay:STARt? (@<chanlist>)
Specifies what starts the over-current protection delay timer.
SCHange starts the over-current delay timer at the END of a settings change in voltage, current, or
output state; allowing for additional protection delay time. Transitions into constant current mode are
ignored DURING a programmed settings change in voltage, current, or output state.
CCTRans starts the over-current delay timer by any transition of the output into current limit mode.
Parameter
Typical Return
SCHange|CCTRans, *RST SCHange
SCH or CCTR
Selects the CCTRans as the current protection delay mode:
CURR:PROT:DEL:STAR CCTR, (@1)
[SOURce:]CURRent:PROTection:STATe 0|OFF|1|ON, (@<chanlist>)
[SOURce:]CURRent:PROTection:STATe? (@<chanlist>)
Enables or disables the over-current protection. If the over-current protection function is enabled and
the output goes into constant current operation, the output is disabled and the Questionable
Condition status register OCP bit is set.
Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
Enable the current protection state: CURR:PROT:STAT ON, (@1)
l
The current limit setting determines when the output goes into constant current operation.
l
An over-current condition can be cleared with
after the cause of the
condition is removed.
[SOURce:]CURRent:RANGe <
value
>, (@<chanlist>)
[SOURce:]CURRent:RANGe? (@<chanlist>)
Sets the output current range on models that have multiple ranges. The value that you enter must be
the highest value in amperes that you expect to source. The instrument selects the range with the
best resolution for the value entered.
Parameter
Typical Return
0 - maximum|MIN|MAX, *RST MAX
The maximum value is 102% of the current rating of
the power module.
<current range>
Sets the output current range to 3 A: CURR:RANG 3, (@1)
7 SCPI Programming Reference
274
Keysight N6705C Operating and Service Guide
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......