l
At the end of the user-defined Arb, the output state of the unit depends upon the
[SOURce;]ARB:TERMinate:LAST program settings.
l
The order in which the values are entered determines the sequence of execution.
l
To create a valid user-defined Arb, the current level, voltage level, BOST, and dwell lists must
either all be the same length, or have a length of 1, which is interpolated as having the same
length as the maximum length list.
[SOURce:]ARB:CURRent:UDEFined:DWELl:POINts? (@<chanlist>)
[SOURce:]ARB:VOLTage:UDEFined:DWELl:POINts? (@<chanlist>)
Returns the number of dwell points in the user-defined Arb, not the point values.
[SOURce:]ARB:SEQuence:STEP:CURRent:UDEFined:DWELl:POINts? <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:UDEFined:DWELl:POINts? <
step#
>, (@<chanlist>)
Returns the number of dwell points in the user-defined Arb for Arb sequence steps.
Parameter
Typical Return
(none)
<number of points>
Step 0 - 99, *RST 0
(none)
Returns the number of current dwell points: ARB:CURR:UDEF:DWEL:POIN? (@1)
Returns the number of current dwell points: ARB:VOLT:UDEF:DWEL:POIN? (@1)
Returns the number of dwell points for step 0: ARB:SEQ:STEP:CURR:UDEF:DWEL:POIN? 0,(@1)
[SOURce:]ARB:CURRent:UDEFined:LEVel <
value
>{<,
value
>},(@<chanlist>)
[SOURce:]ARB:CURRent:UDEFined:LEVel? (@<chanlist>)
[SOURce:]ARB:VOLTage:UDEFined:LEVel <
value
>{<,
value
>},(@<chanlist>)
[SOURce:]ARB:VOLTage:UDEFined:LEVel? (@<chanlist>)
Specifies the level of each current or voltage point in a user-defined Arb. A comma-delimited list of up
to 511 points may be programmed.
[SOURce:]ARB:SEQuence:STEP:CURRent:UDEFined:LEVel <
value
>{<,
value
>}, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:UDEFined:LEVel? <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:UDEFined:LEVel <
value
>{<,
value
>}, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:UDEFined:LEVel? <
step#
>, (@<chanlist>)
Specifies the level of each current or voltage point in a user-defined Arb for Arb sequence steps. The
last value in the sequence is the step number.
Keysight N6705C Operating and Service Guide
261
7 SCPI Programming Reference
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......