Parameter
Typical Return
(none)
<number of points>
Step 0 - 99, *RST 0
(none)
Returns the number of current BOST points: ARB:CURR:UDEF:BOST:POIN? (@1)
Returns the number of voltage BOST points: ARB:VOLT:UDEF:BOST:POIN? (@1)
Returns the number of BOST points for step 0: ARB:SEQ:STEP:CURR:UDEF:BOST:POIN? 0,(@1)
[SOURce:]ARB:CURRent:UDEFined:DWELl <
value
>{<,
value
>},(@<chanlist>)
[SOURce:]ARB:CURRent:UDEFined:DWELl? (@<chanlist>)
[SOURce:]ARB:VOLTage:UDEFined:DWELl <
value
>{<,
value
>},(@<chanlist>)
[SOURce:]ARB:VOLTage:UDEFined:DWELl? (@<chanlist>)
Specifies the dwell time for each user-defined current Arb point. A comma-delimited list of up to 511
points may be programmed. Dwell time is the time that the output will remain at a specific point. Dwell
times can be programmed from 0 through 262.144 seconds with the resolution shown in the table
below:
[SOURce:]ARB:SEQuence:STEP:CURRent:UDEFined:DWELl <
value
>{<,
value
>}, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:UDEFined:DWELl? <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:UDEFined:DWELl <
value
>{<,
value
>}, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:UDEFined:DWELl? <
step#
>, (@<chanlist>)
Specifies the dwell time for each user-defined current Arb point for Arb Sequence steps. The last value
in the sequence is the step number.
Range in seconds
Dwell resolution
Range in seconds
Dwell resolution
All Models except N678xA
Models N678xA only
0 - 0.262144
1 microsecond
0 - 0.268435
1.024 microsecond
0 - 2.62144
10 microseconds
0 - 2.68435
10.24 microsecond
0 - 26.2144
100 microseconds
0 - 26.8435
102.4 microsecond
0 - 262.144
1 millisecond
0 - 268.435
1.024 millisecond
Parameter
Typical Return
0 - 262.144 (all except N678xA SMU)
0 - 268.435 (N678xA SMU only
*RST 1 point with a value of 0.001
<dwell time>
Step 0 - 99, *RST 0
(none)
Program a current dwell list: ARB:CURR:UDEF:DWEL 0.1,0.2,0.3,0.4,0.5,(@1)
Program a voltage dwell list: ARB:VOLT:UDEF:DWEL 0.1,0.2,0.3,0.4,0.5,(@1)
Program a dwell list for step 0: ARB:SEQ:STEP:CURR:UDEF:DWEL 0.1,0.2,0.3,0.4,0.5, 0,(@1)
7 SCPI Programming Reference
260
Keysight N6705C Operating and Service Guide
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......