[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:STARt[:LEVel] <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:STARt[:LEVel]? [MIN|MAX,] <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:STARt[:LEVel] <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:STARt[:LEVel]? [MIN|MAX,] <
step#
>, (@<chanlist>)
Specifies the current or voltage before and after the pulse occurs for Arb sequence steps.
Parameter
Typical Return
Current: 0 - maximum|MIN|MAX, *RST MIN
Voltage: 0 - maximum|MIN|MAX, *RST MIN
The maximum value varies, depending on the current or voltage
rating of the power module.
<start value>
Step 0 - 99, *RST 0
(none)
Programs a start- and end-level current: ARB:CURR:PULS:STAR 1,(@1)
Programs a start- and end-level voltage: ARB:VOLT:PULS:STAR 1,(@1)
Programs a start- and end-level for step 0: ARB:SEQ:STEP:CURR:PULS:STAR 1, 0,(@1)
[SOURce:]ARB:CURRent:PULSe:STARt:TIMe <
value
>, (@<chanlist>)
[SOURce:]ARB:CURRent:PULSe:STARt:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:STARt:TIMe<
value
>, (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:STARt:TIMe? [MIN|MAX,] (@<chanlist>)
Specifies the delay in seconds after the trigger is received, but before the pulse occurs. Referenced to
t0 in the Pulse diagrams.
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:STARt:TIMe <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:STARt:TIMe? [MIN|MAX,] <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:STARt:TIMe <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:STARt:TIMe? [MIN|MAX,] <
step#
>, (@<chanlist>)
Specifies the delay in seconds after the trigger is received, but before the pulse occurs for Arb
sequence steps.
Parameter
Typical Return
0 – 262.144 | MIN | MAX, *RST 0
<start time>
Step 0 - 99, *RST 0
(none)
Programs a start time for the current pulse: ARB:CURR:PULS:STAR:TIM 1,(@1)
Programs a start time for the voltage pulse: ARB:VOLT:PULS:STAR:TIM 1,(@1)
Programs a start time for step 0: ARB:SEQ:STEP:CURR:PULS:STAR:TIM 1, 0,(@1)
7 SCPI Programming Reference
242
Keysight N6705C Operating and Service Guide
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......