![Keysight Technologies N6705C Скачать руководство пользователя страница 119](http://html.mh-extra.com/html/keysight-technologies/n6705c/n6705c_operating-and-service-manual_1962048119.webp)
Arb Sequence Parameters
Parameter
Description
Step <n>
Each step of the sequence includes a step number, an arbitrary waveform, step time, and pacing option. The
total number of steps determines the length. Use the
▲ ▼
navigation keys to scroll through the steps.
Name
The name of the Arb. Select an Arb from the dropdown list. Select Edit or Properties to edit the waveform. The
Pulse waveform edit fields are shown above.
Time
Displays the time that is allocated to the step in the Edit function. This time does not include the number of
repeat counts
Pacing
Displays the pacing of the step. Dwell-paced transitions to the next step when the dwell time has elapsed. Trig-
ger-paced transitions to the next step when an external trigger is received. If the step time completes before the
trigger occurs, the step remains at the last Arb value while waiting for the trigger.
Add
Inserts a step below the selected step; values are copied from the previous step.
Delete
Deletes the presently selected step.
Edit
Edits the selected Arb. The Pulse waveform edit fields are shown above. Move to the next step specifies the step
pacing. Repeat count specifies how many times the Arb repeats. Continuous can only be selected if the Arb is
trigger-paced.
Up
Move up through the steps.
Down
Move down through the steps.
Load (.seq format)
Loads a previously created sequence file.
Save (.seq format)
Saves the present voltage or current sequence.
Total time
Indicates the total run-time of the sequence.
Waveform Quality
Indicates the number of points allocated to the continuously changing portion of the following waveforms:
sine, trapezoid, ramp, and exponential. The more of these waveforms that are added to the sequence, the fewer
points will be allocated, down to a minimum of 16 points.
Advanced
Lets you edit the common properties that apply to the entire sequence. See
for more
information.
Keysight N6705C Operating and Service Guide
119
3 Using the Source Functions
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......