Service Information
2-
Troubleshooting the Test Set
For assistance refer to
“Keysight Support, Services, and Assistance”
If the test set is not operating properly, use the following procedures to aid in
isolating the problem. Refer
to the
“Test Set Diagrams and Graphics”
starting on
Troubleshooting, DC Level
Power Supply and Fans
1.
Verify the front panel Standby switch is operational.
a.
The rear panel fan and front panel DC Power LEDs should be on
when the Standby switch is in the on position.
2.
AC Line voltage checks.
a.
Ensure the proper AC line voltage is present at the instrument line
cord.
b.
Remove the AC power cord from the instrument. Check the
instrument AC line module
c.
Remove the instrument bottom cover. Near the rear panel are two
fuse holders, remove the 8 amp fuses and verify that they are
operational. See
.
Keysight N5261/62A User’s & Service Guide
63
3.
Ex
ternal DC Power checks.
Verify DC voltage levels on each of the front panel DC power
connectors. Pins 1 & 3 = +12 V (± 0.8 V), Pins 4 & 6 = ground
1
3
6
4
7
Pin 1 (+12V)
Pin 6 (gnd)
Pin 3 (+12V)
Pin 4 (gnd)
a.
Figure 2-1
2
Front Panel M
illimeter
Module DC Power Connector
Содержание N5261A
Страница 1: ...User s and Service Guide Keysight Technologies N5261A and N5262A Millimeter Head Controller...
Страница 6: ......
Страница 7: ...Keysight N5261 62A User s Service Guide 1 Installation System Operation...
Страница 51: ...Keysight N5261 62A User s Service Guide 45 Installation System Operation Figure 1 30 Uncalibrated S12 and S21...
Страница 54: ...48 Keysight N5261 62A User s Service Guide...
Страница 55: ...Keysight N5261 62A User s Service Guide 49 Service Information...
Страница 63: ...Service Information 2 Figure 2 8 Sloped Attenuator Cable Assembly Keysight N5261 62A User s Service Guide 57...