8-12 Keysight N5247B Service Guide
General Purpose Maintenance Procedures
Keysight License Manager
8-
Enabling or Removing Options
There are two types of options:
— Hardware: Hardware options involve adding additional hardware to the
analyzer. After the proper hardware has been installed in the analyzer, the
appropriate license can be installed using the Keysight License Manager.
It is necessary to re-enable all installed hardware options, if the midplane
board is replaced.
— Software: Software options add features or functionality to the analyzer
without the need for additional hardware. These options are enabled by
installing the appropriate licenses using the Keysight License Manager
(KLM). Refer to
https://www.keysight.com/us/en/assets/9018-04534/installation-guides
(N5242-90024).
To enable or remove an option:
1.
Start the Keysight Software Manager. Refer to
.
2.
In the Keysight License Manager window that opens, in the left hand
column, press
local host
>
Local Licenses
. A list of available options,
similar to the list below, will appear. Refer to
S93015A/B - Dynamic uncertainty for S-parameters
S93025A/B - Basic pulsed-RF measurements
S93086A/B - Gain Compression Application
S93087A/B- IMD Measurements
029 - Full Noise Measurements
219 - Src/Rcvr Atten & Bias Ts 2-Port
419 - Src/Rcvr Atten & Bias Ts 4-Port
224 - 2nd Src w/Combiner & Switches
423 - Combiner & Switches
3.
Press and hold (or right click with a mouse) the option that you wish to
either enable or remove, and then click
Install
or
Delete
, whichever is
appropriate.
Some applications require a license key file that is provided by Keysight. If
you do not have the required license key file, contact Keysight for
“Contacting Keysight” on page 2-7
.
Содержание N5247B
Страница 10: ...10 Keysight N5247B Service Guide Contents ...
Страница 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Страница 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Страница 227: ...Keysight N5247B Service Guide 5 57 Theory of Operation Receiver Group Operation Figure 5 18 4 Port Receiver Group ...
Страница 228: ...5 58 Keysight N5247B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 19 2 Port Receiver Group ...
Страница 238: ...5 68 Keysight N5247B Service Guide Theory of Operation Noise Measurement Group Operation Option 029 5 ...
Страница 243: ...Keysight N5247B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure ...
Страница 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Страница 381: ...Keysight N5247B Service Guide 6 143 Replaceable Parts Replaceable Parts Listings This page intentionally left blank ...
Страница 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 548: ...7 104 Keysight N5247B Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 565: ...Installation Note Xxxxx xxxxx 3 ...