7-60 Keysight N5247B Service Guide
Repair and Replacement Procedures
Removing and Replacing the A55 Solid State Drive (SSD)
7-
Removing and Replacing the A55 Solid State Drive (SSD)
Certain unique files exist on the SSD that are necessary for proper operation of
your analyzer. These files must be copied to another location to allow them to
be installed onto the new SSD after it has been installed.
If you are replacing the SSD, the following procedure must be performed first.
Copy Unique Files from the Solid State Drive
— If installing an SSD for Windows XP: if the user has loaded unique
calibration kit information, navigate to
C:\Program
Files\Keysight\Network Analyzer
and copy USER_CALKITFILE to a
USB flash memory drive. Also copy any personal user files that you wish to
preserve.
Obsolete. Reference only.
— If installing an SSD for Windows 7: for more information on the N8984A
SSD, refer to the
Windows 7 Operating System Upgrade Kit Installation
Note
https://www.keysight.com/us/en/assets/9018-04311/installation-guides
/9018-04311.pdf
(N8984-90001).
Obsolete. Reference only.
— If installing an SSD for Windows 10: for more information on the N8985A
SSD, refer to the
Windows 10 Operating System Upgrade Kit Installation
Note
https://www.keysight.com/us/en/assets/9018-04733/installation-guides
/9018-04733.pdf
(N8985-90001).
Tools Required
— T-10 TORX driver (set to 9 in-lb; for solid state drive replacement)
— ESD grounding wrist strap
Removal Procedure
Refer to
for this procedure.
1.
Disconnect the analyzer power cord.
2.
Position the analyzer for access to the rear panel.
3.
Loosen the SSD thumb screw (item
).
4.
Pull the SSD out from the CPU assembly, using the handle (item
).
Содержание N5247B
Страница 10: ...10 Keysight N5247B Service Guide Contents ...
Страница 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Страница 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Страница 227: ...Keysight N5247B Service Guide 5 57 Theory of Operation Receiver Group Operation Figure 5 18 4 Port Receiver Group ...
Страница 228: ...5 58 Keysight N5247B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 19 2 Port Receiver Group ...
Страница 238: ...5 68 Keysight N5247B Service Guide Theory of Operation Noise Measurement Group Operation Option 029 5 ...
Страница 243: ...Keysight N5247B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure ...
Страница 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Страница 381: ...Keysight N5247B Service Guide 6 143 Replaceable Parts Replaceable Parts Listings This page intentionally left blank ...
Страница 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 548: ...7 104 Keysight N5247B Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 565: ...Installation Note Xxxxx xxxxx 3 ...