Keysight N5224A/25A Service Guide 6-3
Replaceable Parts
Assembly Replacement Sequence
Assembly Replacement Sequence
The following steps describe how to replace an assembly in the network
analyzer.
Step 1.
Identify the faulty group. Begin with
Follow up
with the appropriate troubleshooting chapter that identifies the faulty
assembly.
Step 2.
Order a replacement assembly. Refer to this chapter.
Step 3.
Replace the faulty assembly and determine what adjustments are necessary.
Chapter 7, “Repair and Replacement Procedures.”
Step 4.
Perform the necessary adjustments. Refer to
Step 5.
Perform the necessary performance tests. Refer to
Step 6.
to review where the calibration
stickers should be placed on the PNA.