4-40 Keysight N5224A/25A Service Guide
Troubleshooting
Measurement System Troubleshooting
4-
3.
Remove the outer cover from the analyzer. Refer to
4.
With the covers off, plug in the analyzer and turn on the power.
Checking the Output Power of the A, B, C, and D Signals
Using a power meter, you can measure the outputs of the A, B, C, and D
signals from the front panel. The measurement results will help you isolate a
faulty assembly. The output of the R receiver cannot be measured because it
would necessitate breaking the phase lock loop, causing all of the signals to be
lost.
Equipment Used for This Check
Equipment Setup
1.
Before starting these checks, zero and calibrate the power meter. (See the
power meter user’s guide for instructions on setting the calibration factor.)
2.
If the
Receiver Display
) is not on the analyzer
screen, perform the following: Press UTILITY
, then
, then
, then
.
3.
Set the sweep speed for a 10 second sweep: Press STIMULUS
,
then
. Set the time to 10.000 seconds in the
Sweep
Time
box.
Procedures described in this document are performed with power
supplied to the product while protective covers are removed. Energy
available at many points may, if contacted, result in personal injury.
Do not operate the analyzer with the outer cover removed for more than
30 minutes, as this could cause the analyzer to overheat which could
result in costly damage.
Equipment Type
Model or
Part Number
Alternate Model or
Part Number
Power meter
E4418B/E4419B
E4418A/E4419A
Power sensor, 2.4 mm
E8487A
None
Adapter, 2.4 mm (f) to 2.4 mm (f)
11900B
85056-60007
System
Service
Utilities
Receiver Display
Sweep
Sweep Time
Содержание N5224A
Страница 124: ...4 52 Keysight N5224A 25A Service Guide Troubleshooting Instrument Block Diagrams Sheet 2 4 ...
Страница 132: ...5 8 Keysight N5224A 25A Service Guide Theory of Operation Synthesized Source Group Operation 5 Figure 5 3 Source Group ...
Страница 149: ...Keysight N5224A 25A Service Guide 5 25 Theory of Operation Receiver Group Operation Figure 5 6 4 Port Receiver Group ...
Страница 150: ...5 26 Keysight N5224A 25A Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 7 2 Port Receiver Group ...
Страница 266: ...6 110 Keysight N5224A 25A Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 346: ...8 18 Keysight N5224A 25A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...