Keysight N5290/1A Service Guide 3-41
System Tests and Adjustments
N5290/1A Millimeter-wave Test System Operator’s Check
N5290/1A Millimeter-wave Test System Operator’s Check
Performing the N5290/1 A System Operator's Check
The N5290/1 A System Operator's Check is done to verify the system to isolate
the problem instrument or after the N5292A test set has been repaired and the
N5292A test Operator's Check has been completed. Refer to
Functional Verification Procedure” on page 3-16
The operator's check is a software driven test that checks the basic operation
of the assemblies in all of the measurement port signal paths. By performing
the operator's check, the following are tested:
— attenuation ranges of all installed attenuators
— calibration of the receivers
— frequency response of the receivers
— phase lock and leveling
— basic functional test of noise floor and trace noise
Accessories Used in the Operator's Check
“Required Service Test Equipment” on page 2-3
STOP!
This section assumes you have completed the N5292A test set
Operator's Check. If you have not completed the N5292A test set
Operator's Check, refer to
“N5292A Functional Verification Procedure” on
.
IMPORTANT!
Before you begin any testing on a System or instrument,
always inspect the connectors on the instrument, the calibration kit,
cables and adapters.
To achieve the maximum system stability, allow the analyzer to warm up
for at least 15 minutes before performing the Operator's Check.
Equipment Type
Part Number
Female short, 1. 0 mm
(any short from the 85059B calibration kits)
Female open, 1. 0 mm
(any open from the 85059B calibration kits)
Содержание N5222B
Страница 3: ...3 ...
Страница 4: ...4 ...
Страница 10: ...6 Keysight N5292A User s and Service Guide Contents ...
Страница 28: ...1 18 Keysight N5290 1A Service Guide Safety and Regulatory Information Site Preparation 1 ...
Страница 86: ...3 50 Keysight N5290 1A Service Guide System Tests and Adjustments Adjustments 3 ...
Страница 128: ...5 10 Keysight N5290 1A Service Guide Theory of Operation Theory of Operation 5 ...
Страница 185: ...Keysight N5290 1A Service Guide 7 9 Repair and Replacement Procedures Removing the Cover Figure 7 3 Outer Cover Removal ...
Страница 214: ...7 38 Keysight N5290 1A Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 224: ...8 10 Keysight N5290 1A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...
Страница 225: ...Installation Note Xxxxx xxxxx 3 ...