Making Measurements
Keysight
N4374B Lightwave Component Analyzer, User’s Guide
77
Measuring Opto-Electrical (OE) Devices
Always perform an electrical calibration of the network
analyzer or load an existing calibration set before you make a
measurement.
For standard measurements, see “Electrical calibration for
single-ended measurements” on page 75.
Before you make your first opto-electrical measurement, you
must initialise the OE measurement. You can see which
measurement is currently initialised at the top left of the trace
screen.
If you have already initialised the OE measurement, you can
connect your device under test and make the measurement by
pressing
or
.
Please refer to the user guide and online help for the network
analyzer for further information.
Initializing the OE measurement
1 Press .
2 Select the “OE Measurement” tab.
3 Select the wavelength at which you are testing.
If you want to use external laser sources, select the optional
External Input, and enter the wavelength of the external
source.